The change of haemoglobin during blood donation, and an assessment of a photometrical method for non-invasive haemoglobin analysis


In Sweden, lowest acceptable haemoglobin levels in blood donators are 125g/L for women and 135g/L for men for a test sample taken in the beginning of the blood donation. Levels, which are 10g/L lower, are accepted if the sample is taken after the blood donation. Earlier studies show that the haemoglobin level decreases for a person that is lying down.The two aims of this study were to examine how much the haemoglobin levels change during blood donation and to examine if the photometrical instrument Pronto-7TM shows equivalent results to that of the established method Cell-Dyn Sapphire.In the study, 120 blood donors participated. Blood samples were taken in the beginning and in the end of the donation. Analyses by Pronto-7TM were done before and after the donation.The haemoglobin level decreased significantly during the blood donation. The difference was in mean value 5,9g/L according to Cell-Dyn Sapphire. The decrease was significantly less than 10g/L. The Pronto-7TM gave levels that were 1,6g/L higher than Cell-Dyn Sapphire in mean and the standard deviation was higher for Pronto-7TM than for Cell-Dyn Sapphire.In conclusion, the decrease of the haemoglobin levels was significantly less than the expected difference 10g/L. Pronto-7TM gives results that differs a little from the results of the established method.